\begin{table}%t1 %\centering \par \caption {\label{tab:logic}Calibration test list. } \begin{tabular}{l l} % centered columns (4 columns) \hline\hline \noalign{\smallskip} T{\sc est id} & D{\sc escription} \\ \hline \multicolumn{2}{c}{F\sc unctionality} \\ {\tt RCA\_{\rm A}MB} & Functional test at ambient \\ {\tt RCA\_CRY} & Functional test at cryo \\ \multicolumn{2}{c}{T\sc uning} \\ {\tt RCA\_TUN} & Gain and offset tuning of the DAE \\ & Tuning of the front end module \\ & (phase switches and gate voltages) \\ \multicolumn{2}{c}{B{\sc asic} P{\sc roperties}} \\ {\tt RCA\_OFT} & Radiometer offset \\ & measurement \\ {\tt RCA\_TNG} & Noise temperature and \\ & photometric gain \\ {\tt RCA\_LIS} & Radiometer linearity \\ \multicolumn{2}{c}{N{\sc oise} P{\sc roperties}} \\ \par {\tt RCA\_STn} & Noise performances tests: \\ & WN, $f_{\rm k}$ and $\alpha$, $\beta$, $r$ \\ {\tt RCA\_UNC} & Verification of the effect \\ & of the radiometer switching \\ & on the noise spectrum \\ \multicolumn{2}{c}{B{\sc and} P{\sc ass} R{\sc esponse}} \\ {\tt RCA\_SPR} & Bandpass \\ \multicolumn{2}{c}{S\sc usceptibility} \\ {\tt RCA\_THF} & Susceptibility to \\ & FEM temperature variations \\ {\tt RCA\_THB} & Susceptibility to \\ & BEM Temperature variations \\ {\tt RCA\_THV} & Susceptibility to V-groove \\ & temperature variations \\ \hline %inserts single line \end{tabular} \tablefoot {The first column reports the test identification. In the second column the purpose of each test is described; WN is the white noise level; $f_{\rm k}$, $\alpha$ are the $1/f$ knee frequency and slope respectively; $\beta$ is the equivalent radiometer bandwidth derived from noise; $r$ is the modulation factor. Apart from the first test {\tt RCA\_{\rm A}MB}, which is performed at ambient temperature, all the other tests are performed at the operational temperature (i.e. at a temperature as close as possible to in-orbit conditions). } \end{table}