Licence Creative CommonsOpen Access article, published by EDP Sciences, under the terms of the Creative Commons Attribution License (https://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.

This article is published in open access under the Subscribe to Open model. Subscribe to A&A to support open access publication.

1 Introduction

The launch of the X-ray Imaging and Spectroscopy Mission (XRISM; Tashiro et al. 2024) has ushered in the era of nondispersive X-ray spectroscopy thanks to its micro-calorimeter Resolve (Ishisaki 2025; Kelley 2025). XRISM will enable unprecedented measurements of fundamental physical quantities in hot plasma, in a temperature and density regime where X-rays are the most accurate probe of its physical conditions (Martizzi et al. 2019).

While high-resolution spectroscopy enables diagnostic measurements of individual emission lines with little dependence on the underlying continuum or the spectral shapes at different energies, spectral analysis over broad energy ranges remains a key tool for providing accurate estimates of plasma physical parameters, especially when observing complex absorption line systems with small equivalent widths. In this context, XRISM/Resolve spectra benefit from concurrent observations with other X-ray observatories, whose scientific payload extends the energy range and spatial coverage - typically at the price of a moderate energy resolution and/or lower photon collecting area. In order for multi-instrument analysis to produce accurate results - or, at least, accurate estimates of the systematic uncertainties - a good understanding of the instrument cross-calibration is crucial. It has been shown that uncertainties in the flux and spectral shape determination from different instruments can constitute an important contribution to the uncertainty budget in some specific science cases (Schellenberger et al. 2015).

In this paper, we present a systematic comparison between the spectra obtained by the XRISM scientific payload - including Xtend, a charged-coupled device (CCD), large field-of-view instrument (Noda et al. 2025; Uchida et al. 2025) - and spectra obtained by NuSTAR (Harrison et al. 2013), the XMM-Newton European Photon Imaging Camera (EPIC; Turner et al. 2001; Strüder et al. 2001) and Reflection Grating Spectrometer (RGS; de Vries et al. 2015), and the Chandra High-Energy Transmission Grating Spectrometer (HETGS; Canizares et al. 2005) during a multi-instrument observational campaign on the Seyfert Galaxy NGC 3783 in July 2024. The paper is organised as follows. We describe the observational campaign in Sect. 2. The data handling and reduction are described in Sects. 3 and 4. Section 5 presents the cross-calibration analysis, including a novel methodology for comparing spectral predictions from different instruments and its application to the NGC 3783 campaign. We summarise our main findings and recommendations in Sect. 6.

2 The NGC 3783 July 2024 observational campaign

The core of our campaign occurred between July 17 and 29, 2024. Coordinated observations with Chandra, the Hubble Space Telescope (HST), NICER, NuSTAR, Swift, XMM-Newton, and XRISM were carried out. The timeline of these observations is shown graphically in Fig. 1. In addition to the observations shown, there were several individual Swift and NICER snapshots preceding our main campaign, and a short (10 ks) Chandra observation in December 2023.

In the present paper, we focus on the cross-calibration of the main X-ray spectroscopic instruments (XRISM, XMM-Newton, Chandra, and NuSTAR). More details about the optical/UV instruments (HST, the OM camera of XMM-Newton, and Swift UVOT) as well as the X-ray monitoring instruments (Swift XRT and NICER) will be given in subsequent papers where the data from these instruments are used. A summary of the observations discussed in this paper is given in Table 1.

The X-ray spectrum of NGC 3783 is very complex and is discussed in Mehdipour et al. (2025, Paper I) and subsequent papers. Relevant to the content of this paper, we only note a strong narrow Fe-K emission line near 6.35 keV (observed energy) and a deep Unresolved Transition Array (UTA), a blend of M-shell ions of iron between 0.7 and 0.8 keV (see for example Mehdipour et al. 2017).

3 Data handling

In this section, we describe general procedures that we applied to the data of all instruments. Data reduction details for individual instruments will be discussed in Sect. 4. Spectral fitting was performed using SPEX software version 3.08.02 (Kaastra et al. 1996).

3.1 Orbital corrections to the energy scale

For the high-resolution instruments (RGS, HETGS, and Resolve), small corrections on the energy scale must be made to account for the motion of the satellite around Earth and the motion of Earth relative to the Solar System barycentre (SSB). For RGS, this is taken into account by default in the analysis software, and for HETGS and Resolve, we determined these corrections based on the ephemerid data for the satellites.

The velocity corrections vSSB are calculated as the velocity of the satellite relative to the SSB, along the line of sight towards NGC 3783. For our campaign, these velocities are all negative, resulting in an apparent additional redshift of the spectrum (the source appears to be moving away from us).

For individual Chandra observations, the mean velocity vSSB was in the range of −24 to −22 km/s, with an rms spread of only 0.07 km/s. It is dominated by the motion of Earth, since Chandra is most of the time far away from Earth with a low relative speed. The mean velocity is −22.98 km/s, and it is sufficiently accurate to apply this correction to all Chandra spectra.

For each XRISM orbit around Earth, the rms spread is 3.7 km/s, due to the higher speed of the satellite in its low-Earth orbit. For individual orbits, the average velocity ranges between −26 and −22 km/s. The mean velocity averaged over all orbits is −22.83 km/s.

One of the strongest absorption lines in the Chandra spectrum is the Si XIV line near 2 keV. It can be measured with an accuracy of 90 km/s in the stacked spectrum. Since we typically observe about nine lines of similar strength in the full spectrum, astrophysical velocity components can be determined with accuracies of about 90/√9 = 30 km/s. This justifies the use of a single and uniform velocity correction of −22.98 km/s for the Chandra spectra. For Resolve, the strongest line in the spectrum is the Fe II Kα emission line near 6.4 keV. It can be determined with a statistical uncertainty of 16 km/s in the stacked spectrum, also justifying the use of a uniform velocity correction vSSB of −22.83 km/s. Technically, we applied these velocity corrections by multiplying the model energy grid values in the response matrix by a factor of (1 + vSSB/c) where c is the speed of light (note the sign!).

thumbnail Fig. 1

Overview of the observations. Horizontal lines indicate the time intervals when the instruments were observing NGC 3783. ‘COS’ indicates the Cosmic Origin Spectrograph on board HST. XRISM contains the Xtend and Resolve instruments, XMM-Newton the pn, mos1, mos2, RGS1, and RGS2 instruments; for Chandra, the HETGS was the only operational instrument.

Table 1

Observations used in this paper.

3.2 Statistics

We used C-statistics instead of χ2-statistics throughout this work. The C-statistic can be used for statistical tests of the model (Kaastra 2017) and, contrary to the χ2 statistic, it does not give a biased flux estimate (Hitomi Collaboration 2017).

Chandra spectra are usually provided with Gehrels errors (Gehrels 1986), but we replaced these by the Poissonian uncertainties equal to the square root of the number of counts. This is important for both the application of the C-statistic and the optimal re-binning; re-binning Gehrels errors would lead to over-estimating the true uncertainties.

Uncertainties on parameters are reported in this paper as root-mean-square uncertainties (68% confidence for a normal distribution). Separate positive- and negative-side uncertainties are only provided when their absolute values differ significantly.

3.3 Background treatment and selection

The C-statistic has to be applied to a model for the sum of the source and background counts in the source extraction region. A model for the source counts is often based on an astrophysical model. It is not always straightforward to obtain a model for the background counts in the source extraction region. For that reason, sometimes a scaled background spectrum derived from an off-source region is subtracted. This background subtracted spectrum does not follow a Poissonian distribution, which is a fundamental requirement for the proper application of the C-statistic. Sometimes an analytical model for the background can be constructed, but in other cases the background is too complex. We constructed for all instruments except Resolve the background model from the observed background spectrum in an offset region by applying Wiener filtering to it (Press et al. 1992). This reduces most of the noise in the background. For the RGS spectra this method is applied to the individual CCDs of each spectrometer (either are active for each of the two RGS detectors). For all other instruments a single filtering procedure is sufficient. For Resolve we used a smoothed version of the non-X-ray background model instead.

The determination of the background for low-Earth orbit satellites depends on the criteria used to exclude time intervals of high-particle flux in the South Atlantic Anomaly. For XRISM, we used the limits recommended in the standard data reduction pipeline for each instrument. We investigated the effect of different time screening criteria or background extraction region on the net background-subtracted NuSTAR spectra. Applying a stricter South Atlantic Anomaly screening to the NuSTAR data1, the net exposure time is reduced by 6% with only an overall 0.8 ± 1.3% flux difference without significant energy dependence.

3.4 Data binning

For all instruments, we applied optimal binning to the spectra and response matrices (Kaastra & Bleeker 2016). The algorithm for obtaining the optimal grid for both the photon energy grid and the observed data channel grid is described in that paper. We used the new rbin option of the SPEX package to produce reduced size spectra and matrices. By optimally binning the data, no information is lost but the size of the matrices and thereby the time needed to convolve a model spectrum with the response matrix are strongly reduced. This is in particular important for the Resolve data, where the process reduces the size of the extralarge response matrix (see Sect. 4.1) by a factor of ~100 (see Table 2).

Table 2

Number of data channels and response elements in the original and final (optimally binned) response matrix for the main instruments involved in our campaign.

4 Data reduction

4.1 XRISM/Resolve

NGC 3783 was observed starting on July 18, 2024, for a total Resolve duration of 451 ks (ObsID = 300050010). Since the Resolve aperture door (‘gate valve’) was closed during this observation, a 252 μm-thick beryllium window attenuates X-ray photons in the soft X-ray band, reducing the effective area (Midooka et al. 2020). This restricted the bandpass to energies above 1.7 keV. The aperture door also has a stainless steel protective screen that reduces the effective area by an additional 28% across the waveband.

The data were processed using XRISM pre-pipeline software version 004_007.20Jun2024_Build8.012. The data reduction was performed using XRISM pipeline processing version 03.00.012.008 and calibration database (CALDB) version 9 (20240815 release). This CALDB release includes initial updates to the Resolve energy scale and line-spread function (LSF) files based on in-orbit calibration data. The temporal gain-drift of the Resolve instrument was corrected at 6 keV using data collected during Earth occultation from 55Fe sources on the Resolve filter wheel, following the approach reported in Porter et al. (2025), see also XRISM Collaboration (2024).

The energy scale error at 6 keV due to intermittent gain monitoring can be estimated using the calibration pixel (pixel 12), a pixel offset from the main array that receives Mn Kα photons continuously from a small, collimated 55Fe source. This estimate is obtained by calculating the energy offset (‘fit shift’) at Mn Kα on the calibration pixel during the main observation period, not including the time period of the gain monitoring. This value is reported by the XRISM Software Data Center in a ‘Resolve Energy Scale Quality Report’ for each observation2. For the NGC 3783 observation this value is ~0.16 eV. To estimate the total uncertainty on the energy scale across the waveband, this error must be combined with the systematic energy scale errors reported in Eckart et al. (2025) of ~ ± 0.3 eV in the 5.4-8.0 keV band, up to ±1.0 eV for energies below 5.4 keV, and up to ±2 eV at energies above 9 keV. The broadband energy scale errors are uncorrelated with the temporal gain-drift correction and can be root-sum-squared (see Sect. 5.3 of Eckart et al. 2025), for a total energy scale uncertainty of ±0.34 eV from 5.4-9 keV, ±1.0 eV for energies below 5.4 keV, and ±2 eV at energies above 9 keV.

Additional screening and energy-dependent rise time cuts were applied according to Mochizuki et al. (2025), following the procedure in the XRISM Quick Start Guide Version 2.13. We also removed pixel 27, an edge pixel on the Resolve detector array, from our analysis. Pixel 27 has been shown to exhibit gain jumps of a few eV on timescales of a few hours, between periods of gain monitoring, and thus current guidance recommends excluding pixel 27. After this screening, the total exposure time was slightly reduced to 439 ks. Only ‘high-resolution primary’ events are used for the spectral analysis. The non-X-ray background spectrum was modelled from the Resolve night-Earth data database4, while the sky background was ignored because its contribution above 2 keV is negligible.

To account for detector redistribution, a redistribution matrix file was created using the rslmkrmf task with the cleaned event file and the CalDB file. This CalDB file was created using a combination of model parameters from ground-calibration measurements and updates from initial on-orbit calibration (Leutenegger 2025). By using the extra-large option with rslmkrmf, the resulting response matrix file (RMF) incorporates all known features of the LSF model: a Gaussian core, low-energy exponential tail, escape peaks, electron-loss continuum, and Si K fluorescence (Eckart et al. 2018; Leutenegger 2025). During the NGC 3783 Resolve observation, the full width at half maximum (FWHM) of the composite Mn Kα spectrum of all array pixels for the calibration monitoring intervals was 4.48 ± 0.01 eV. The predicted FWHM resolution (core LSF) as a function of energy is provided in the CalDB file for each pixel. The uncertainty on the FWHM energy resolution (core LSF) is estimated to be 50 meV at 2 keV, 100 meV at 5 keV, and 200 meV at 8 keV for high-resolution events (Leutenegger 2025). An auxiliary response file was generated using the xaarfgen task, assuming a point-like source at the aim point as input.

4.2 XRISM/Xtend

The Xtend instrument operated in 1/8-window mode to reduce frame time and mitigate potential pileup. The latest CALDB version 20241115 has been utilised in the data reduction. After applying standard event screening and filtering out periods containing high particle background, we obtained an effective exposure of 372 ks. The spectrum was extracted from the central 2′ × 5.5′ region, while the background spectrum was obtained from an adjacent area of the same size. The background region exhibits a flat light curve, indicating no strong contamination from any solar wind events. The redistribution matrix and auxiliary response files were generated using the xtdrmf and xaarfgen routines.

4.3 NuSTAR

The NuSTAR data were extracted using the NuSTAR pipeline version 0.4.9 with the calibration database of July 29, 2024. For the spectral extraction we chose a circular region with radius 120". For the background region we chose an annulus around the source with inner and outer radii of 180" and 230", respectively, to minimise the effects of possible gradients of the background over the detector area.

4.4 XMM-Newton/EPIC

All EPIC data were taken with the thin filter and in small window mode. EPIC source (plus background) and background spectra were extracted from calibrated and concatenated event lists. They were generated with the standard data reduction pipelines e[mp]proc and the latest calibration files available at the time the data reduction was performed (September 2024) using Science Analysis System (SAS) version 21.0.0 (Gabriel et al. 2004). Source spectra were extracted from annular regions centred around the optical active galactic nucleus (AGN) position, with an outer radius of 40" and 42", and removing a core of the point spread function (PSF) corresponding to 15" and 20" for the EPIC-pn and EPIC-MOS, respectively, to mitigate mild pileup in the strongest flux states of the source during the campaign (manifesting for EPIC-pn count rates >19 s-1 in the 0.3-7 keV energy band). Background spectra were extracted from circular regions of radii 45"-50" and 120"-140" for EPIC-pn and EPICMOS, respectively, after ensuring that the background regions do not include any serendipitous X-ray source and that the EPIC-pn background regions are at the same height in detector coordinate as the source regions to ensure that the same correction for the charge transfer inefficiency applies. Observation-specific instrument responses were generated with the standard SAS command arfgen and rmfgen.

4.5 XMM-Newton/RGS

RGS spectra were extracted using the XMM-Newton SAS version 21.0.0 with the following command: rgsproc withsrc=true srcra=174.757342 srcdec=-37.738670 withrectification=yes. This last parameter applies an empirical correction (‘rectification’) to the RGS effective area response aiming at aligning it to that of EPIC-pn. This correction is the default as of version 22.0.0 of SAS.

4.6 Chandra/HETGS

Chandra data were extracted using the standard scripts of version 4.16 of the Chandra Interactive Analysis of Observations (CIAO) package. We executed the Chandra_repro process, and combined positive and negative spectral orders using combine_grating_spectra.

5 Cross-calibration results

5.1 Effects contributing to cross-calibration uncertainty

No instrument has a perfect calibration, and the instruments involved in our campaign are no exception. Discrepancies become apparent when the spectra from two or more instruments are compared. In general, the following factors can contribute to a mismatch between an observed spectrum and the predicted spectrum from a given astrophysical model: uncertainties in (a) the effective area, (b) the background modelling, (c) pileup effects, (d) the energy scale calibration, (e) the spectral redistribution function, and (f) time coverage. We briefly consider these effects (b-f) in the following subsections. Effective area (a) is discussed in detail in Sect. 5.4.

5.1.1 Background modelling

Because NGC 3783 is a relatively bright source, the modelled background is in general small, and therefore uncertainties in the background estimate do not significantly affect the derived source spectrum. The only exception may be the higherenergy parts of the NuSTAR spectrum, where due to slight non-uniformity over the detector image the choice of background region has some effects (see Sect. 4.3). Alternatively, we also extracted the background from a circular region with radius 130" sufficiently far away from the source region, a common practice for observers. This region yields a 1.6% higher source flux below 30 keV and 4% higher source flux above that energy. In our further analysis, we used the annular region centred around the source for the background.

5.1.2 Pileup and implications of the mitigation strategy

Pileup or count rate limits are not a problem for XRISM Resolve or for Xtend with the observing mode that we used. Only for the XMM-Newton EPIC cameras there is some pileup in the cores of the detector image of the source (see Sect. 4.4). Pileup affects the determination of the fluxes as well as the spectral shape due to pattern migration. To mitigate pileup, EPIC spectra were therefore generated from annular regions, after excising the core of the PSF primarily affected by pileup. This procedure ideally produces spectra that are fully representative of those extracted from a full aperture. However, this is not necessarily the case. Uncertainties on the encircled energy fraction calculated over the wings of the PSF may constitute the dominant - and, regrettably, poorly known - source of calibration uncertainties. For this reason, we postulate that the EPIC spectra of the NGC 3783 2024 observational campaign are affected by an unknown effective area calibration error. Therefore, we will not discuss the EPIC cameras in the pairwise cross-calibration analysis discussed in Sect. 5.4.

5.1.3 Energy-scale uncertainties

Energy-scale uncertainties have no strong effect on highresolution spectra, apart from a small bias in derived redshifts of lines. For lower-resolution instruments like CCDs, gain uncertainties can have a significant impact on the cross-calibration. As an example, we fitted the Xtend spectrum of NGC 3783 with a spline model (see Sect. 5.2), then applied an artificial gain shift of 2% to the model spectrum (using the reds component of SPEX), and computed the residuals with respect to that redshifted model. Due to the strong Fe-K emission line, a large ±10% oscillation of the fit residuals appears close to this line with a significance of 18σ. It also leads to a ~5% large-scale deviation near 3 keV, due to the combined curvature of the spectrum and effective area of the instrument. Therefore, attention must be paid to even small uncertainties in the energy scale of moderate-resolution instruments when assessing the cross-calibration with other instruments.

5.1.4 Spectral redistribution tails

Uncertainties in the spectral redistribution function for highresolution instruments do not result in major problems, because the bulk of the photons are registered in the narrow core of the redistribution function and only a small fraction (<~5%) appear in a near-core, low-energy tail. The situation is different for medium-resolution instruments like CCDs, in particular at low energies. Here a substantial fraction of all photons can be observed in strong, low-energy tails redistributed from the core, difficult to disentangle from photons at the bona fide correct energy. This implies that uncertainties in the low-energy effective area and the redistribution function are strongly entangled.

This holds in particular for the EPIC cameras of XMM-Newton and the Xtend detector of XRISM.

5.1.5 Variability and time coverage

Flux-dependent changes in sensitivity are not expected for our instruments. A possible complication for our analysis is the timevariability of the source. For a proper comparison, strict simultaneity is mandatory. Unfortunately, the time interval where all instruments are simultaneously operating is too small to be useful for cross-calibration purposes. Therefore, we relied on a pair-wise comparison of instruments, taking for each pair the longest strictly simultaneous time interval. The effect of time variability on some of this pair-wise analysis is discussed in Sect. 5.5. For extracting spectra for a pair of instruments, their individual good time intervals (GTIs) are combined by the LHEASOFT task ftmgtime using only intervals where both instruments were on.

Some instruments (like the RGS on XMM-Newton) make separate GTIs for each CCD readout sequence. When they are combined with those from another instrument that follows the same procedure, there may be no GTI left that has 100% coverage with the other instrument, resulting in a net zero exposure time. We mitigated this by making longer GTI intervals for those instruments by first merging adjacent GTIs that are separated by less than the CCD readout sequence (a few seconds).

Another issue is related to the orbits of the low-Earth satellites XRISM and NuSTAR. The orbits cause interruptions in the light curve every 1.6 hour but with different phases, meaning only 30% of the available NuSTAR exposure time coincides with the available XRISM exposure. Fortunately, the flux variations of NGC 3783 are not extreme. To demonstrate this, we used the power spectrum of Markowitz (2005) as used by Li et al. (2023) to simulate the light curve. For 6500 realisations of 80 consecutive orbits (which is what we have for NuSTAR), the scatter in the ratio of the flux during the time NuSTAR was operating (50% of the time) and the mean flux between the start and end of the NuSTAR observation is only 0.5% around the mean value of 1. We conclude that for our purpose we can use safely the full NuS-TAR dataset rather than the limited data with full overlap with XRISM.

5.2 Methodology

Given the above-mentioned caveats, we compared the two instruments as follows. We first performed a spline fit to the higher-resolution spectrum in the pair. For this, we used the spin model of SPEX, with a linear energy grid (for Resolve) or a linear wavelength grid (for the grating spectrometers). For the resolution of the spline, we chose a number close to the resolution of the instrument, but it is optimised for any specific instrument pair.

A great advantage of this method is that it is agnostic with respect to the complex astrophysical details of the spectrum. The best-fit C-stat for this model is often significantly smaller than the expected value for a realistic astrophysical model, because the spline also tries to accommodate for the Poissonian noise of the spectrum. We carried out simulations using this spline method for model spectra with known shapes and found that the reconstructed spectrum with the spline model does not produce any additional bias in the source flux and thus can be used reliably.

After obtaining this best-fit model to the highest-resolution spectrum of the pair, we folded the lower-resolution instrument through this model and inspected the relative fit residuals. Rebinning the residuals by factors of typically 10 to 100 allows us to see the large-scale discrepancies of the fluxes measured by the two instruments. Based on these findings, we derived simple analytical correction factors for one instrument to get it in agreement with the other instrument, thereby making it possible to obtain broad-band spectral fits with astrophysical models by combining the two datasets. In doing that, we assumed that flux differences are solely or primarily determined by uncertainties in the calibration of the effective area. Details will be given in Sect. 5.4.

thumbnail Fig. 2

Relative residuals (data/model-1) against a common spline model applied to RGS, Resolve, and NuSTAR data (details in the text). Differences between residuals mainly indicate differences of the effective area with respect to the effective area calibrations. The residuals have been binned for clarity. The black and grey residuals refer to RGS1/2 below 3 keV and to FPMA/B above 3 keV.

5.3 Global view

As a first step, we made a joint fit of two high-resolution instruments (XMM-Newton RGS and XRISM Resolve) in the band below 12 keV together with the only hard X-ray instrument (NuS-TAR) above 12 keV using all available data. For the model, we chose three splines covering together the full energy range of these instruments, with an energy grid spacing of approximately twice the bin size of these instruments: a 326-point linear wavelength grid between 5.5 and 38 Å in the RGS and lowest-energy Resolve band, a 975-point linear energy grid between 5.5 Å (approximately 2.25 keV) and 12 keV in the Resolve band, and a 67-point linear energy grid between 12 and 78 keV at higher energies. We determined the residuals for each instrument relative to this model, plus the computed residuals for the instruments that were not used for the fit: XRISM Xtend and XMM-Newton pn, MOS1, and MOS2 (see Fig. 2).

The main differences in the relative residuals are due to calibration uncertainties; the effects of variability (different time intervals for the instruments) are only a few percent. From low to high energy, we encounter the following features. In the 12 keV band the RGS residuals appear to be 10-15% below those of Chandra and XRISM. The EPIC (pn and MOS) residuals are about 20% below those of XRISM around 3 keV, and above that energy MOS and pn start deviating strongly from each other. The NuSTAR fluxes in the 3-12 keV band are within a few percent of those measured by XRISM. The apparent small jump near 12 keV in the NuSTAR residuals is an artefact caused by using exclusively Resolve and RGS below 12 keV and NuSTAR above this energy.

Observers often try to compensate these kinds of residuals by adjusting the effective areas by a constant, which is a free parameter for each instruments. However, we see from Fig. 2 that the residuals are strongly energy-dependent. In order to better understand the origin of these differences and to fix them, we now turn to a detailed pair-wise comparison.

thumbnail Fig. 3

Residuals of the Xtend spectrum relative to the Resolve model (black). The red line is a simple analytical representation of these residuals. The dotted purple lines indicate the 1σ combined statistical uncertainty of the Resolve spectra of NGC 3783, binned to the energy grid of the Xtend data. They represent a systematic uncertainty of the spline model for Resolve.

5.4 Pair-wise instrument comparison

5.4.1 Resolve-Xtend

We produced Resolve and Xtend spectra using their common GTIs (348 ks exposure time). The Resolve spectrum was fit using a 900-point spline plus a Gaussian component for the narrow Fe-K emission line. It appears that the centroid of the Fe-K line in the Xtend spectrum is shifted by 15 eV relative to Resolve and the instrumental Si-line by 30 eV. We corrected the Xtend data here and in the rest of the paper using linear interpolation of the energy scale with the gain model of SPEX with two parameters p0 and p 1 as indicated in Fig. 3. This figure shows the remaining residuals of the Xtend spectrum relative to the Resolve model.

In the 4-8 keV energy band the two instruments agree apart from a ~2.5% offset. The narrow peak at 4 keV is not statistically significant; it is due to a 2σ dip in the Resolve model that is enhanced in Xtend due to the three times higher effective area. The deficit of 10% near 2 keV is likely due to uncertainty in the Resolve effective area; see Sect. 5.4.2. There is a small excess up to ~10% at 12 keV. This may be caused by the slightly larger size of the CCD depletion region in Xtend compared with the current response model. We approximated the excess by a simple analytical expression, as shown in the inset of Fig. 3.

5.4.2 Reassessing the Resolve throughput at low energies

The deficit in the Xtend spectrum near 2 keV (Fig. 3) can alternatively be interpreted as an excess in the flux predicted by the Resolve spectrum. A similar excess near 2 keV is observed in the Resolve data for 3C 273, which has a much simpler intrinsic spectrum in this energy range (E. Miller, private communication). Also, removal of this excess would bring the Resolve flux into closer agreement with the fluxes measured by all XMM-Newton instruments (RGS and EPIC) at this energy.

We developed a simple analytical expression to adjust the Resolve flux to better match that of Xtend in this energy band. As shown in Fig. 4, the analytical expression (red curve) approximates the transmission curve computed with SPEX (black curve). The black curve represents the transmission of a layer containing 1024 Be atoms per m2, equivalent to approximately 7.5 μm of solid beryllium. The analytical approximation was crafted for convenience, given the rapid energy-dependent variation in the transmission of the thick beryllium window of the Resolve gate valve over this waveband. This adjustment is not intended to definitively attribute the flux discrepancy between Resolve and Xtend to calibration errors or uniformity issues in the beryllium window. However, we note that the gate valve’s ground calibration was performed only at energies above 2.6 keV (Midooka et al. 2020), leaving the possibility that the model extrapolation to lower energies introduced error. Future measurements at a synchrotron facility using the spare unit may lead to refinement of the gate valve transmission curve at lower energies.

thumbnail Fig. 4

Simple analytical approximation (red) used to reduce excess Resolve flux compared to Xtend and other CCD instruments at the low-energy end of the Resolve band. This expression was derived by approximating the transmission of 1024 atoms m2 (black).

5.4.3 Resolve-NuSTAR

Due to the different orbital phases of XRISM and NuSTAR, the strictly simultaneous joint exposure time is 70 ks. We show in Sect. 5.1.5 that ignoring these orbital gaps leads to differences smaller than 0.5%. For this reason, we compared the full NuS-TAR spectrum (226 ks) with the full Resolve spectrum between the start and end of the NuSTAR observation. The Resolve spectrum was fitted using a spline between 2 and 12 keV with 801 grid points (spacing 12.5 eV) and a narrow Gaussian Fe-K line (C-stat 2475, expected 3199±80; see Kaastra 2017).

Figure 5 shows a comparison of the NuSTAR focal plane module A (FPMA) spectrum with this model for the Resolve spectrum. At higher energies, FPMA shows an excess of about 5%; near 6 keV the two instruments agree, but at lower energies the residual is +4%. We modelled this empirically using the formula shown in the figure.

thumbnail Fig. 5

Residuals of the FPMA spectrum relative to the Resolve model (black). The red line is a simple analytical representation of these residuals.

5.4.4 FPMA-FPMB

We fitted a spline model to the full FPMA spectrum and investigated the residuals for FPMB. These are higher by only 0.4 ± 0.2%.

5.4.5 RGS

Figure 2 shows a strong deficit of about 15% of the RGS and EPIC instruments of XMM-Newton near 2 keV, relative to most other instruments. Motivated by this, we considered the effective area calibration of the RGS. The refinement of the RGS calibration is still ongoing and too complex to describe in full detail here, so we focus on an important intermediate product.

Figure 6 shows the stacked residuals of all RGS spectra collected until 2019 on the blazar Mkn 421. All individual spectra have been modelled by a broken power law with Galactic absorption. The effective area includes all default corrections except for the pn-rectification factor (de Vries et al. 2015).

The figure shows that the residuals decrease rapidly for wavelengths shorter than ~12 Å. Near 6 Å or 2 keV they are about 10% below the extrapolation of the longer-wavelength spectrum. While the cause of this decrease has not yet been explained, we made an empirical correction factor to account for it as indicated in the figures.

With this correction taken into account, we made a new joint spline fit through the RGS, Resolve, and NuSTAR spectra, taking into account the effective area corrections for these instruments obtained so far. These results will be discussed in more detail later, but they show an additional discrepancy between RGS1 and RGS2 around 7 Å (Fig. 7). As discussed below, this discrepancy can be attributed to RGS2. Therefore, we excluded RGS2 between 6.55 and 7.55 Å in the fit.

Figure 7 shows that around 7 Å, RGS2 deviates up to ~23% from RGS1, Xtend and HETGS medium-energy grating (MEG). A broad-band fit with a realistic astrophysical model through the combined Resolve and RGS spectra (Zhao et al., in prep.) confirms that RGS1 represents the source spectrum properly near 7 Å. The RGS calibration becomes more uncertain for wavelengths below 7 Å, and in particular near chip gaps. Such a gap occurs for RGS2 between CCDs 1 and 2 near 7.2 Å. For this reason, we applied the empirical correction shown in Fig. 7 to the RGS2 data.

thumbnail Fig. 6

Stacked residuals of Mrk 421 RGS1 and RGS2 spectra as described in the text. Here x represents the wavelength in Å, and y the data/model ratio. The red curves show a simple analytical model for these residuals, with shape and parameters shown in red in the insets. Negative outliers are associated with the hot or bad pixels that are not perfectly taken into account in the simplified stacking procedure.

5.4.6 RGS-Xtend

After incorporating the corrections to the RGS, Resolve and NuSTAR spectra presented above, we folded the Xtend spectrum through this model (Fig. 8). In addition, we made a spline fit to the RGS spectrum of another radio-loud active AGN, 3C273. This source is observed every year in the framework of an international cross-calibration campaign involving all operational X-ray observatories. The RGS and Xtend data were reduced as specified in Sects. 4.5 and 4.2, respectively. We folded that model through the simultaneous Xtend data of that source, which were analysed in the same way as the NGC 3783 data. At least for energies larger than about 0.8 keV, the residuals are similar to those shown in Fig. 8 and therefore point to a common discrepancy between the effective area of RGS versus Xtend. A detailed analysis of the 3C 273 data will be presented in a separate paper (Miller et al., in prep.).

thumbnail Fig. 7

Residuals for a joint fit with a spline model through the RGS1 (black), RGS2 (grey, excluding wavelength between 6.55 and 7.55 Å in the fit), Resolve and NuSTAR (not shown) instruments. For comparison we show the residuals to this model for Xtend (purple histogram) and MEG (orange); these last two instruments were not included in the fit. The red line is a fit to the RGS2 residuals, with model and parameters indicated in the inset.

thumbnail Fig. 8

Relative residuals of the Xtend spectrum of NGC 3783 relative to a spline fit to the simultaneous RGS, Resolve, and NuSTAR spectra. The dotted purple lines indicate the 1σ combined statistical uncertainty of the combined RGS+Resolve spectra of NGC 3783, binned to the energy grid of the Xtend data. They represent a systematic uncertainty of the spline model for NGC 3783.

5.4.7 Resolve-Chandra HETGS

We fitted the Resolve spectrum between 1.8 and 6.7 keV using a spline with 981 knots (spacing 5 eV). The Resolve spectrum was modified according to the empirical correction of the soft X-ray throughput described in Sect. 5.4.2, which affects the low-energy part of the spectrum. Due to the small effective area at low energies the fluxed Resolve spectrum becomes uncertain, as shown by the blue lines in Fig. 9.

thumbnail Fig. 9

Relative residuals of the Chandra HETGS-MEG spectrum (binned by a factor of 10) relative to the Resolve model (black). The red line is a simple analytical representation of these residuals. Blue lines indicate the ±1σ uncertainty on the Resolve spectrum binned to the MEG bins.

thumbnail Fig. 10

Relative residuals of a spline fit to the Chandra MEG spectrum, and the HEG spectrum relative to that model.

5.4.8 Chandra HEG and MEG

A comparison of the two gratings of the HETGS can be made using the full dataset, because the two operate simultaneously. We fitted the MEG spectrum with a spline and folded the high-energy grating (HEG) spectrum through it. The ratio is ≃-1% for λ ≤ -5 Å, and ≃-4% for λ ≥ -5 Å (Fig. 10).

5.5 Corrections for source variability for joint instrument analysis

The ultimate goal of this paper requires production of spectra that can be analysed together. In particular when the time-averaged spectrum is considered, corrections must be made for the different time intervals when the instruments were operating. Fortunately, the time interval during which Xtend was operated encompasses the operational time intervals of the other instruments. Therefore, for each instrument, we can determine the ratio of the 0.3-12 keV band Xtend flux when that other instrument was operating, against the flux measured during the full Xtend exposure. The major contribution to this ratio is due to the variability of the primary AGN power-law component. Variations on smaller energy scales cannot be fully excluded using the medium-resolution Xtend data, and would need special care in variability studies using the high-resolution instruments involved, but we expect their impact to be small on the X-ray broadband continuum.

Figure 11 shows the ratio y of the Xtend spectra between the full exposure and the common time interval when RGS was also being operated. It can be approximated by a simple power law. Similar results were found for the Chandra and NuSTAR data. Best-fit power-law parameters A and b (y = AEb with E in keV) are given in Table 3. The deviations from unity are significant but small, except for the Chandra data, and will be taken into account in the analysis of the combined dataset in future papers.

For Chandra, five of the seven observations have been taken while XRISM was also observing NGC 3783. The ratio of this Chandra spectrum to the full Chandra spectrum can be approximated by 1 + 0.07/λ1.2 with λ the wavelength of a photon in Å.

thumbnail Fig. 11

Ratio of the Xtend spectrum during the time when RGS was observing to the full Xtend spectrum. The red curve is a simple powerlaw approximation to this ratio.

Table 3

Parameters A and b for power-law fits to the ratio of Xtend spectra extracted during a time interval common to another instrument and during the full exposure.

Table 4

Energy-dependent multiplication factors to be applied to the effective area for different instruments.

thumbnail Fig. 12

Relative residuals of the spectra. The residuals of RGS and MEG have been binned by a factor of 50, for Resolve a factor of 200, NuSTAR a factor of 5 below 18 keV and 20 above 18 keV, and a factor of 5 for Xtend.

6 Discussion and conclusions

Based on our partially simultaneous data from multiple instruments and considerations of the cross-calibration of these instruments, we plan to make adjustments to selected instrument effective area curves in our subsequent papers, which will discuss multi-instrument data taken during the NGC 3783 2024 observational campaign. Relative to the standard calibration available at the time of this paper, corrections will be applied to the low- or high-energy side of the effective area curves, as summarised in Table 4. The factor f (E) listed in the table represents the multiplier to the effective area.

Because the data are not strictly simultaneous, additional corrections need to be made to compensate for missing data. This is facilitated by the fact that AGNs like NGC 3783 often show a strong correlation between the photon index and flux of the dominant power-law component, at least in the 2-10 keV band. We determined these correction factors by comparing the full Xtend spectra with Xtend spectra during the time that the other instrument was observing NGC 3783. The correction factors are well represented by power-law corrections (Table 3).

We verified these correction factors by making a spline fit to the combined RGS, Resolve, andNuSTAR data (the NuSTAR data were restricted to the band above 10 keV in the fit). We show the residuals relative to this model for the full bands of these instruments and for Xtend and the Chandra MEG in Fig. 12. While the adjusted Resolve, NuSTAR, and RGS spectra agree within 12% with this model, showing the effectiveness of our procedure, the Chandra MEG spectrum still shows some fluctuations up to ~±6% in the 1-3 keV band. This is likely due to the relatively small overlap between the XRISM and Chandra data. As previously noted, two of the seven Chandra spectra were taken before or after the XRISM observations, and the other five overlap with only about 10% of the duration of the XRISM observation. In fact, most of the overlapping Chandra exposure is near the start or the end of the XRISM observation, epochs characterised by relatively high flux. Thus, the discrepancies may be caused by spectral variations in the astrophysical source. This will be explored in forthcoming papers.

The Xtend spectra below 2 keV still show differences of up to 14% relative to the model, which is dominated here by the RGS instrument. This may be related to the complexity of the calibration of CCD detectors at low energies, or the effective area calibration of the RGS, but more effort, beyond the scope of this paper, is needed to test that hypothesis.

We plan to apply the empirical corrections described in Tables 3 and 4 to subsequent papers, which will present the multi-instrument analysis of the observational campaign of NGC 3783. These corrections are not based on a physical model of the telescope or the instruments. We therefore discourage the reader from interpreting the results in this paper as ‘the crosscalibration status’ between the XRISM scientific payload and other operational X-ray observatories. Effective area calibration may subtly depend on source properties and observational configuration. A full understanding of the typical differences in the determination of flux and spectral shape between XRISM and other observatories can only be derived from an analysis of the full sample of sources included in the XRISM in-flight calibration plan (Miller et al. 2020). Furthermore, our understanding of the XRISM payload calibration is still in its initial stages. Changes in the effective area of the Resolve and Xtend instruments are likely to be encoded in future versions of the calibration database. The status we describe in this paper - and in particular through Fig. 2 - constitutes an encouragement to carefully take cross-calibration uncertainties into account before undertaking multi-instrument global fits involving the XRISM scientific instruments.

Acknowledgements

We are grateful to the mission planners of all missions involved for their efforts in scheduling and re-scheduling all our coordinated observations as well as possible, which is not a trivial task. We thank Lucien Kuiper (SRON) for providing us with the velocity corrections to the Solar System Barycentre. SRON is supported financially by NWO, the Netherlands organization for scientific research. Part of this work was performed under the auspices of the U.S. Department of Energy by Lawrence Livermore National Laboratory under Contract DE-AC52-07NA27344. The material is based upon work supported by NASA under award number 80GSFC21M0002. This work was supported by JSPS KAKENHI grant numbers JP22H00158, JP22H01268, JP22K03624, JP23H04899, JP21K13963, JP24K00638, JP24K17105, JP21K13958, JP21H01095, JP23K20850, JP24H00253, JP21K03615, JP24K00677, JP20K14491, JP23H00151, JP19K21884, JP20H01947, JP20KK0071, JP23K20239, JP24K00672, JP24K17104, JP24K17093, JP20K04009, JP21H04493, JP20H01946, JP23K13154, JP19K14762, JP20H05857, and JP23K03459. This work was supported by NASA grant numbers 80NSSC20K0733, 80NSSC18K0978, 80NSSC20K0883, 80NSSC20K0737, 80NSSC24K0678, 80NSSC18K1684, and 80NNSC22K1922. LC acknowledges support from NSF award 2205918. CD acknowledges support from STFC through grant ST/T000244/1. L. Gallo acknowledges financial support from Canadian Space Agency grant 18XARMSTMA. MS acknowledges the support by the RIKEN Pioneering Project Evolution of Matter in the Universe (r-EMU) and Rikkyo University Special Fund for Research (Rikkyo SFR). AT and the present research are in part supported by the Kagoshima University postdoctoral research program (KU-DREAM). SY acknowledges support by the RIKEN SPDR Program. IZ acknowledges partial support from the Alfred P. Sloan Foundation through the Sloan Research Fellowship. This work was supported by the JSPS Core-to-Core Program, JPJSCCA20220002. The material is based on work supported by the Strategic Research Center of Saitama University. M. Mehdipour acknowledges support from NASA through the XRISM Guest Scientist (XGS) grant 80NSSC23K0995, a grant for HST program number 17273 from the Space Telescope Science Institute (operated by the Association of Universities for Research in Astronomy, Inc., under NASA contract NAS5-26555), and NuSTAR grant 80NSSC25K7126. We deeply regret that our co-author Katja Pottschmidt passed away on 17 June 2025, the same day that this paper was submitted.

References

  1. Canizares, C. R., Davis, J. E., Dewey, D., et al. 2005, PASP, 117, 1144 [NASA ADS] [CrossRef] [Google Scholar]
  2. de Vries, C. P., den Herder, J. W., Gabriel, C., et al. 2015, A&A, 573, A128 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  3. Eckart, M. E., Adams, J. S., Boyce, K. R., et al. 2018, J. Astron. Telesc. Instrum. Syst., 4, 021406 [Google Scholar]
  4. Eckart, M. E., Brown, G. V., Chiao, M. P., et al. 2025, J. Astron. Telesc. Instrum. Syst., 11 [Google Scholar]
  5. Gabriel, C., Denby, M., Fyfe, D. J., et al. 2004, in Astronomical Society of the Pacific Conference Series, 314, Astronomical Data Analysis Software and Systems (ADASS) XIII, eds. F. Ochsenbein, M. G. Allen, & D. Egret, 759 [Google Scholar]
  6. Gehrels, N. 1986, ApJ, 303, 336 [Google Scholar]
  7. Harrison, F. A., Craig, W. W., Christensen, F. E., et al. 2013, ApJ, 770, 103 [Google Scholar]
  8. Hitomi Collaboration (Aharonian, F., et al.) 2017, Nature, 551, 478 [Google Scholar]
  9. Ishisaki, Y. e. a. 2025, J. Astron. Telesc. Instrum. Syst., 11 [Google Scholar]
  10. Kaastra, J. S. 2017, A&A, 605, A51 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  11. Kaastra, J. S., & Bleeker, J. A. M. 2016, A&A, 587, A151 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  12. Kaastra, J. S., Mewe, R., & Nieuwenhuijzen, H. 1996, in UV and X-ray Spectroscopy of Astrophysical and Laboratory Plasmas, eds. K. Yamashita & T. Watanabe, 411 [Google Scholar]
  13. Kelley, R. L. e. a. 2025, J. Astron. Telesc. Instrum. Syst., 11 [Google Scholar]
  14. Leutenegger, M. A. e. a. 2025, J. Astron. Telesc. Instrum. Syst., 11 [Google Scholar]
  15. Li, C., Kaastra, J. S., Gu, L., & Mehdipour, M. 2023, A&A, 680, A44 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  16. Markowitz, A. 2005, ApJ, 635, 180 [NASA ADS] [CrossRef] [Google Scholar]
  17. Martizzi, D., Vogelsberger, M., Artale, M. C., et al. 2019, MNRAS, 486, 3766 [Google Scholar]
  18. Mehdipour, M., Kaastra, J. S., Kriss, G. A., et al. 2017, A&A, 607, A28 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  19. Mehdipour, M., Kaastra, J. S., Eckart, M. E., et al. 2025, A&A, 699, A228 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  20. Midooka, T., Tsujimoto, M., Kitamoto, S., et al. 2020, SPIE Conf. Ser., 11444, 114445C [Google Scholar]
  21. Miller, E. D., Sawada, M., Guainazzi, M., et al. 2020, SPIE Conf. Ser., 11444, 1144426 [Google Scholar]
  22. Mochizuki, Y., Tsujimoto, M., Kilbourne, C. A., et al. 2025, J. Astron. Telesc. Instrum. Syst., 11, 042002 [Google Scholar]
  23. Noda, H., Mori, K., Tomida, H., et al. 2025, PASJ, in press [arXiv:2502.08030] [Google Scholar]
  24. Porter, F. S., Kilbourne, C. A., Chiao, M., et al. 2025, J. Astron. Telesc. Instrum. Syst., 11 [Google Scholar]
  25. Press, W. H., Teukolsky, S. A., Vetterling, W. T., & Flannery, B. P. 1992, Numerical recipes in FORTRAN. The art of scientific computing [Google Scholar]
  26. Schellenberger, G., Reiprich, T. H., Lovisari, L., Nevalainen, J., & David, L. 2015, A&A, 575, A30 [NASA ADS] [CrossRef] [EDP Sciences] [Google Scholar]
  27. Strüder, L., Briel, U., Dennerl, K., et al. 2001, A&A, 365, L18 [Google Scholar]
  28. Tashiro, M., Watanabe, S., Maejima, H., et al. 2024, SPIE Conf. Ser., 13093, 130931G [Google Scholar]
  29. Turner, M. J. L., Abbey, A., Arnaud, M., et al. 2001, A&A, 365, L27 [CrossRef] [EDP Sciences] [Google Scholar]
  30. Uchida, H., Mori, K., Tomida, H., et al. 2025, PASJ, 77, 523 [Google Scholar]
  31. XRISM Collaboration (Audard, M., et al.) 2024, PASJ, 76, 1186 [Google Scholar]

All Tables

Table 1

Observations used in this paper.

Table 2

Number of data channels and response elements in the original and final (optimally binned) response matrix for the main instruments involved in our campaign.

Table 3

Parameters A and b for power-law fits to the ratio of Xtend spectra extracted during a time interval common to another instrument and during the full exposure.

Table 4

Energy-dependent multiplication factors to be applied to the effective area for different instruments.

All Figures

thumbnail Fig. 1

Overview of the observations. Horizontal lines indicate the time intervals when the instruments were observing NGC 3783. ‘COS’ indicates the Cosmic Origin Spectrograph on board HST. XRISM contains the Xtend and Resolve instruments, XMM-Newton the pn, mos1, mos2, RGS1, and RGS2 instruments; for Chandra, the HETGS was the only operational instrument.

In the text
thumbnail Fig. 2

Relative residuals (data/model-1) against a common spline model applied to RGS, Resolve, and NuSTAR data (details in the text). Differences between residuals mainly indicate differences of the effective area with respect to the effective area calibrations. The residuals have been binned for clarity. The black and grey residuals refer to RGS1/2 below 3 keV and to FPMA/B above 3 keV.

In the text
thumbnail Fig. 3

Residuals of the Xtend spectrum relative to the Resolve model (black). The red line is a simple analytical representation of these residuals. The dotted purple lines indicate the 1σ combined statistical uncertainty of the Resolve spectra of NGC 3783, binned to the energy grid of the Xtend data. They represent a systematic uncertainty of the spline model for Resolve.

In the text
thumbnail Fig. 4

Simple analytical approximation (red) used to reduce excess Resolve flux compared to Xtend and other CCD instruments at the low-energy end of the Resolve band. This expression was derived by approximating the transmission of 1024 atoms m2 (black).

In the text
thumbnail Fig. 5

Residuals of the FPMA spectrum relative to the Resolve model (black). The red line is a simple analytical representation of these residuals.

In the text
thumbnail Fig. 6

Stacked residuals of Mrk 421 RGS1 and RGS2 spectra as described in the text. Here x represents the wavelength in Å, and y the data/model ratio. The red curves show a simple analytical model for these residuals, with shape and parameters shown in red in the insets. Negative outliers are associated with the hot or bad pixels that are not perfectly taken into account in the simplified stacking procedure.

In the text
thumbnail Fig. 7

Residuals for a joint fit with a spline model through the RGS1 (black), RGS2 (grey, excluding wavelength between 6.55 and 7.55 Å in the fit), Resolve and NuSTAR (not shown) instruments. For comparison we show the residuals to this model for Xtend (purple histogram) and MEG (orange); these last two instruments were not included in the fit. The red line is a fit to the RGS2 residuals, with model and parameters indicated in the inset.

In the text
thumbnail Fig. 8

Relative residuals of the Xtend spectrum of NGC 3783 relative to a spline fit to the simultaneous RGS, Resolve, and NuSTAR spectra. The dotted purple lines indicate the 1σ combined statistical uncertainty of the combined RGS+Resolve spectra of NGC 3783, binned to the energy grid of the Xtend data. They represent a systematic uncertainty of the spline model for NGC 3783.

In the text
thumbnail Fig. 9

Relative residuals of the Chandra HETGS-MEG spectrum (binned by a factor of 10) relative to the Resolve model (black). The red line is a simple analytical representation of these residuals. Blue lines indicate the ±1σ uncertainty on the Resolve spectrum binned to the MEG bins.

In the text
thumbnail Fig. 10

Relative residuals of a spline fit to the Chandra MEG spectrum, and the HEG spectrum relative to that model.

In the text
thumbnail Fig. 11

Ratio of the Xtend spectrum during the time when RGS was observing to the full Xtend spectrum. The red curve is a simple powerlaw approximation to this ratio.

In the text
thumbnail Fig. 12

Relative residuals of the spectra. The residuals of RGS and MEG have been binned by a factor of 50, for Resolve a factor of 200, NuSTAR a factor of 5 below 18 keV and 20 above 18 keV, and a factor of 5 for Xtend.

In the text